Specifications
Measurement Principal: Continuous Hydrogen Collection & Analysis
Range (with AT-S Probe): 0 to 2,000 pL/cm2/s
Accuracy: +/-10% Logged Reading +/-Resolution
Response: <3 Minutes
Calibration: Annual; Can Be Checked with Hydrosteel Function Test Kit
Can Be Applied to Steel Surface up to 500 C
Data Log: Every 10 Minutes
Power: 24 Volt DC